首页> 外文期刊>Catheterization and cardiovascular interventions: Official journal of the Society for Cardiac Angiography & Interventions >Changes in coil morphology following transcatheter coil occlusion of the patent ductus arteriosus using a modified snare technique.
【24h】

Changes in coil morphology following transcatheter coil occlusion of the patent ductus arteriosus using a modified snare technique.

机译:使用改良的圈套技术,经导管盘闭动脉导管未闭后,线圈形态发生变化。

获取原文
获取原文并翻译 | 示例
           

摘要

We observed a decrease in length of the Gianturco coils following transcatheter occlusion of the patent ductus arteriosus (PDA). Coil length was measured on chest radiograph within 24 hr of coil placement and compared to the length at the time of follow-up. Echocardiograms were also reviewed for evidence of duct recanalization. Twenty-seven patients met inclusion criteria. The median time to follow-up was 5 months (1--12 months). Twenty-four out of 27 (89%) patients had a decrease in coil length by 1.9 +/- 1.1 mm (P < 0.01). This was an average decrease of 16%. One patient had an increase in length and two patients had no change in coil length. Nineteen out of 27 patients had echocardiograms. Despite the change in coil length, there was no evidence of flow acceleration in the pulmonary artery or descending aorta. No patients had evidence of duct recanalization.
机译:我们观察到动脉导管未闭(PDA)经导管阻塞后Gianturco线圈长度的减少。在放置线圈后24小时内,在胸部X光片上测量线圈长度,并将其与随访时的长度进行比较。还检查了超声心动图,以发现导管再通的证据。二十七名患者符合纳入标准。中位随访时间为5个月(1--12个月)。 27例患者中有24例(89%)的线圈长度减少了1.9 +/- 1.1 mm(P <0.01)。平均下降了16%。一名病人的长度增加,而两名病人的线圈长度没有变化。 27例患者中有19例接受了超声心动图检查。尽管线圈长度有所变化,但没有证据表明肺动脉或降主动脉有血流加速。没有患者有导管再通的证据。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号