首页> 外文期刊>Nuclear physics, B >Energy response and resolution of YAP : Ce matrix for imaging applications in nuclear medicine
【24h】

Energy response and resolution of YAP : Ce matrix for imaging applications in nuclear medicine

机译:YAP的能量响应和分辨率:Ce矩阵在核医学中的成像应用

获取原文
获取原文并翻译 | 示例
           

摘要

In order to design an accurate imaging detector for Nuclear Medicine, the energy performances of several YAP:Ce crystal arrays were investigated upon. Indeed, the new multipillar configuration used for the crystal gamma detector, needs an attentive pillars dimensions study in order to optimize the light yield, the detection efficiency and the spatial linearity. At first, the YAP:Ce matrix were coupled with a EMI 9765 traditional phototube; measurements of light yield and energy linearity were taken by irradiating the crystals with Am-241, Ba-133 and Cs-137 point-like sources. Then, the energy response and resolution of a 40 mm x 40 mm detection area YAP:Ce matrix coupled with a Hamamatsu R2486 Position Sensitive PMT were attentively analyzed. Differences were found in the energy response on the whole detection area, but with an adequate equalization procedure it is possible to recover the response homogeneity as well as improving the detector energy resolution. [References: 9]
机译:为了设计用于核医学的精确成像检测器,研究了几种YAP:Ce晶体阵列的能量性能。的确,用于晶体伽马探测器的新型多柱构造需要进行细心的支柱尺寸研究,以优化光输出,探测效率和空间线性度。首先,将YAP:Ce矩阵与EMI 9765传统光电管耦合;通过用Am-241,Ba-133和Cs-137点状光源照射晶体来测量光产率和能量线性。然后,重点分析了结合Hamamatsu R2486位置敏感PMT的40 mm x 40 mm检测区域YAP:Ce矩阵的能量响应和分辨率。在整个检测区域的能量响应中发现了差异,但是通过适当的均衡程序,可以恢复响应均匀性并提高检测器能量分辨率。 [参考:9]