首页> 外文期刊>Catheterization and cardiovascular interventions: Official journal of the Society for Cardiac Angiography & Interventions >Cobrahead malformation of the Amplatzer septal occluder device: an avoidable compilation of percutaneous ASD closure.
【24h】

Cobrahead malformation of the Amplatzer septal occluder device: an avoidable compilation of percutaneous ASD closure.

机译:Amplatzer隔室封堵器装置的眼镜蛇畸形:可避免的经皮ASD闭合汇编。

获取原文
获取原文并翻译 | 示例
           

摘要

During deployment of an Amplatzer atrial septal occluder device to close a secundum ASD, the right atrial disk and waist of the device twisted, resulting in the cobrahead malformation. Postulated mechanisms for this complication include twisting of the device during loading into the delivery catheter and catching of leading edge of the device on the LA free wall or appendage, causing twisting during deployment. Retrieval of the device into the catheter and even removal of the device from the patient to allow manual untwisting may be required to allow the device to return to its original conformation for successful redeployment.
机译:在部署Amplatzer房间隔封堵器装置以关闭继发性ASD期间,右房间盘和装置的腰部会扭曲,从而导致眼镜蛇头畸形。用于这种并发症的假定机制包括在加载到输送导管中期间装置的扭曲以及在LA自由壁或附件上捕获装置的前缘,从而在展开期间引起扭曲。可能需要将装置收回到导管中,甚至将装置从患者身上取下以允许手动解扭,以使装置恢复到其原始形态以成功进行重新部署。