Safety and efficacy of target vessel catheterization with the new steerable microcatheter direxion compared with a standard microcatheter: A prospective, preclinical trial

机译:与标准微导管相比,采用新型可控微导管定向术的目标血管导管插入术的安全性和有效性:一项前瞻性临床前试验

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To assess the potential of a steerable microcatheter in a comparative preclinical trial. Methods: A total of 100 small target vessels of the lower limbs with a maximum diameter of 3 mm were prospectively randomized to catheterize with either the preshaped torqueable Direxion? (J tip shape; Boston Scientific, Natick, MA) or a similarly steam-shaped Renegade? microcatheter (Boston Scientific) in a porcine model. Catheterization was first performed in combination with a microguidewire and afterwards without. Results: No significant differences were found for the mean vessel diameter in the Direxion (1.53 ± 0.44 mm; n = 50) or Renegade (1.62 ± 0.43 mm; n = 50; p = 0.35) group. Guidewire-assisted catheterization was successful in all target vessels, whereas access was achieved in most cases with the guidewire alone. However, when it became necessary to steer the Direxion actively, this was regarded as key to obtain vessel access in three of four target vessels (75 %). Vessel catheterization without guidewire was significantly more successful with the Direxion (88 %; n = 44) compared with the Renegade (32 %; n = 16; p < 0.0001). In addition, this catheterization technique was also significantly faster with the Direxion compared with guidewire-assisted vessel catheterization with the Renegade (16.1 ± 14.4 sec compared with 27.1 ± 24.7 sec; p = 0.011). Conclusions: The Direxion microcatheter demonstrated unique steerability characteristics, which makes it a promising new tool especially for complex coaxial endovascular procedures.
机译:目的:在临床前比较试验中评估可操纵的微导管的潜力。方法:前瞻性地将总共100个最大直径为3 mm的下肢小目标血管随机分配给预成型的可扭转Direxion?导管。 (J尖头形状;马萨诸塞州内蒂克市的Boston Scientific)还是类似蒸汽形状的叛徒?猪模型中的微型导管(波士顿科学公司)。首先与微导丝结合进行导尿,然后不进行导尿。结果:Direxion(1.53±0.44 mm; n = 50)或Renegade(1.62±0.43 mm; n = 50; p = 0.35)组的平均血管直径无显着差异。导丝辅助的导管插入术在所有目标血管中均成功完成,而在大多数情况下,仅使用导丝即可达到。但是,当有必要主动操纵Direxion时,这被认为是获得四个目标船只中的三个(75%)的船只通道的关键。与Renegade(32%; n = 16; p <0.0001)相比,使用Direxion(88%; n = 44)时,没有导丝的导管插入术显着更成功。此外,与使用Renegade的导丝辅助血管导管术相比,使用Direxion的导管术也明显更快(16.1±14.4秒,而27.1±24.7秒; p = 0.011)。结论:Direxion微导管显示出独特的可操纵性特征,这使其成为有前途的新工具,尤其是对于复杂的同轴血管内手术。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号