首页> 外文期刊>Neuromodulation: journal of the International Neuromodulation Society >Comparison of a Novel Fixation Device With Standard Suturing Methods for Spinal Cord Stimulators
【24h】

Comparison of a Novel Fixation Device With Standard Suturing Methods for Spinal Cord Stimulators

机译:一种新型的固定装置与标准缝合方法的脊髓刺激器的比较

获取原文
获取原文并翻译 | 示例
       

摘要

Introduction: Spinal cord stimulation is a well-established treatment for chronic neuropathic pain of the trunk or limbs. Currently, the standard method of fixation is to affix the leads of the neuromodulation device to soft tissue, fascia or ligament, through the use of manually tying general suture. A novel semiautomated device is proposed that may be advantageous to the current standard. Comparison testing in an excised caprine spine and simulated bench top model was performed. Materials and Methods: Three tests were performed: 1) perpendicular pull from fascia of caprine spine; 2) axial pull from fascia of caprine spine; and 3) axial pull from Mylar film. Six samples of each configuration were tested for each scenario. Standard 2-0 Ethibond was compared with a novel semiautomated device (Anulex fiXate). Upon completion of testing statistical analysis was performed for each scenario. Results: For perpendicular pull in the caprine spine, the failure load for standard suture was 8.95 lbs with a standard deviation of 1.39 whereas for fiXate the load was 15.93 lbs with a standard deviation of 2.09. For axial pull in the caprine spine, the failure load for standard suture was 6.79 lbs with a standard deviation of 1.55 whereas for fiXate the load was 12.31 lbs with a standard deviation of 4.26. For axial pull in Mylar film, the failure load for standard suture was 10.87 lbs with a standard deviation of 1.56 whereas for fiXate the load was 19.54 lbs with a standard deviation of 2.24. Conclusions: These data suggest a novel semiautomated device offers a method of fixation that may be utilized in lieu of standard suturing methods as a means of securing neuromodulation devices. Data suggest the novel semiautomated device in fact may provide a more secure fixation than standard suturing methods.
机译:简介:脊髓刺激是治疗躯干或四肢慢性神经性疼痛的公认方法。当前,固定的标准方法是通过使用手动绑扎普通缝合线将神经调节装置的引线固定到软组织,筋膜或韧带上。提出了可能对当前标准有利的新型半自动化装置。在切除的山羊脊柱和模拟台式模型中进行了对比测试。材料和方法:进行了三个测试:1)从山羊脊柱筋膜垂直拉动; 2)从山羊脊柱筋膜轴向拉动; 3)聚酯薄膜的轴向拉力。针对每种情况测试了每种配置的六个样本。将标准2-0 Ethibond与新型半自动设备(Anulex fiXate)进行了比较。测试完成后,对每种情况进行统计分析。结果:垂直牵引山羊脊柱,标准缝合的失败载荷为8.95磅,标准偏差为1.39,而固定缝合的载荷为15.93磅,标准偏差为2.09。对于轴向牵引山羊脊柱,标准缝合的失败载荷为6.79磅,标准偏差为1.55,而固定载荷为12.31磅,标准偏差为4.26。对于轴向拉动聚脂薄膜,标准缝合的失败载荷为10.87磅,标准偏差为1.56,而固定载荷的载荷为19.54磅,标准偏差为2.24。结论:这些数据表明,一种新型的半自动化装置提供了一种固定方法,可以代替标准缝合方法来固定神经调节装置。数据表明,与标准缝合方法相比,新型半自动化设备实际上可以提供更安全的固定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号