摘要

PURPOSE: Optically guided ultrasound imaging has been used in our department since 2003 in order to implement an on line correction scheme in intensity modulated radiation therapy of prostate carcinoma. PATIENTS AND METHODS: The corrections observed during the initial time period of the system (17 patients) are compared to those observed more recently (10 patients). Treatment margins are calculated. RESULTS: Overall systematic errors decreased between 2003 and 2006, and are presently statistically not different from zero. Random errors remain the same (max 4.3 mm). Proposed margins are 7 mm both in lateral and longitudinal direction and 8.4 mm in anteroposterior. CONCLUSION: Ultrasound can be used for on line correction of both positioning and internal organs motion errors and allows reduction of the margins between clinical and planning volume.
机译:目的:自2003年以来,我们部门一直使用光导超声成像技术,以在前列腺癌的调强放射治疗中实施在线校正方案。患者与方法:将系统初始时间段内观察到的校正(17例患者)与最近观察到的校正(10例患者)进行比较。计算治疗余量。结果:总体系统误差在2003年至2006年之间有所下降,目前统计上与零没有差异。随机误差保持不变(最大4.3毫米)。建议的横向和纵向边距均为7 mm,前后距为8.4 mm。结论:超声可用于定位和内部器官运动误差的在线校正,并可以减少临床和计划量之间的界限。