To study the cause of myocardial injury after elective transvenous insertion of a cardioverter/defibrillator, we measured troponin T (TnT) release in 27 patients. Five patients needing only replacement of the generator with threshold testing served as controls. Compared to the control group, a significant increase in TnT occurred in the patients undergoing insertion, which was greatest in the group receiving screw-in electrodes. There was no correlation between TnT and the number and energy of the shocks and the time in fibrillation. Duration of surgery and TnT release were positively associated. Cardiac injury therefore mainly results from mechanical trauma during insertion of the electrodes and only to a lesser degree from electrical and ischemic lesions.
机译:为了研究选择性经静脉插入心脏复律器/除颤器后心肌损伤的原因,我们测量了27例患者的肌钙蛋白T(TnT)释放。将仅需要通过阈值测试更换发生器的五名患者作为对照。与对照组相比,插入患者的TnT显着增加,在接受旋入电极的组中最大。 TnT与电击的次数和能量以及颤动时间之间没有相关性。手术时间和TnT释放呈正相关。因此,心脏损伤主要由电极插入期间的机械损伤引起,而仅在较小程度上由电和缺血损伤引起。