While microsatellite genotyping has found wide application in many fields, a number of causes that could lead to error in microsatellite genotyping have been previously reported. Here we report another cause of error that we term 'Electrophoresis Artefacts' (EA), which arise when high concentrations of polymerase chain reaction (PCR) products are electro-phoresed in ABI 377 automated sequencers. We describe the phenomenon of EAs, characterize PCR and gel running conditions that cause them and suggest a simple method of overcoming this problem.
机译:尽管微卫星基因分型已在许多领域得到广泛应用,但先前已报道了许多可能导致微卫星基因分型错误的原因。在这里,我们报告了另一个错误原因,我们将其称为“电泳伪像”(EA),当在ABI 377自动测序仪中对高浓度的聚合酶链反应(PCR)产品进行电泳时,就会出现这种错误。我们描述了EA的现象,表征了导致EA的PCR和凝胶运行条件,并提出了克服这一问题的简单方法。