In this work, one type of thermoluminescence dosimeter low density polyethylene grafted by acrylic acid acrylamide, acrylic acid actinamide polymer film (LDPE-g-p (AAm/AAc)) was irradiated by ~(60)Co therapeutic gamma ray dose. This work deals with the case of sulfonated grafted film which has good properties such as thermal stability, and good dosimetric properties. The various sources of uncertainty for this type of thermoluminescence dosimeter under study were analyzed. The uncertainty budget tables for radiation measurements were declared. For the used procedure, these uncertainties multiply the coverage factor equal 2 to obtain the expanded uncertainty at 95% confidence level. The combined uncertainty does not exceed 6.3%. The results suggested that, the expanded uncertainty at 95% confidence level should be added to the value of measurements to obtain the accurate dose.
机译:在这项工作中,用〜(60)Co治疗γ射线辐照了一种由丙烯酸丙烯酰胺接枝的热致发光剂量计低密度聚乙烯,丙烯酸act酰胺聚合物薄膜(LDPE-g-p(AAm / AAc))。这项工作涉及具有良好性能(如热稳定性)和良好的剂量特性的磺化接枝膜的情况。分析了正在研究的这种类型的热释光剂量计的各种不确定性来源。宣布了辐射测量的不确定性预算表。对于所使用的过程,这些不确定性将覆盖因子乘以2,以在95%置信度下获得扩展的不确定性。合并的不确定性不超过6.3%。结果表明,应将95%置信水平下的扩展不确定度添加到测量值中,以获得准确剂量。