A time-harmonic target-field method for designing shielded RF coils in MRI

机译:磁共振成像中屏蔽射频线圈的时谐目标场方法

获取原文
获取原文并翻译 | 示例
           

摘要

An inverse method for designing shielded, high-frequency RF coil systems is presented. A time-harmonic target-field approach is taken to find current densities on primary and shield RF coils. The correspondingly induced magnetic field is required to be homogeneous upon an asymmetrically located target region within the scanner and to be zero over an exterior target region. The finite length of the coil system is represented via a Fourier series expansion for the current density. Some new regularization penalty functions are considered in the solution process. A technique is given for creating coil winding patterns from complex, time-dependent current densities. To demonstrate the power of this design method, several coil windings are displayed for the primary and shield coils along with the corresponding fields.
机译:提出了一种设计屏蔽的高频射频线圈系统的逆方法。采用时谐目标场方法来找到初级和屏蔽RF线圈上的电流密度。要求相应的感应磁场在扫描仪内不对称定位的目标区域上是均匀的,并且在外部目标区域上为零。线圈系统的有限长度通过电流密度的傅立叶级数展开表示。在求解过程中考虑了一些新的正则化罚函数。给出了一种根据复杂的,随时间变化的电流密度创建线圈绕组模式的技术。为了演示这种设计方法的强大功能,将为初级线圈和屏蔽线圈显示几个线圈绕组以及相应的字段。