Application of a motion capture data glove for hand and wrist ergonomic analysis during laparoscopy

机译:运动捕获数据手套在腹腔镜检查过程中手部和腕部人体工程学分析中的应用

获取原文
获取原文并翻译 | 示例
           

摘要

Objectives: This study aims to analyze the surgeons' hand spatial configuration during the use of two different instrument handles for laparoscopy, by obtaining information from the data glove CyberGlove(R), and establishing existing risk levels for wrist disorders. Material and methods: Fifty surgeons participated in this study and were distributed into three groups (novices, intermediate and experts). Each subject carried out suturing and dissection tasks on a physical simulator, using axial handled or ring-handled instruments, respectively. Hand and wrist positions were registered by the CyberGlove(R) and a modified RULA method was applied to establish appropriate risk levels for wrist disorders. Results: We found statistically significant differences in seven of the eleven glove sensors when comparing both tasks. RULA method showed that all subjects, with the exception of the experts using an axial-handled instrument, assume a prejudicial wrist posture during the practice of suturing and dissection tasks on the simulator. Conclusions: Data glove CyberGlove(R) allows for the distinction between two laparoscopic exercises performed with different instruments. Also, laparoscopic intracorporeal suturing when performed with an axial-handled needle holder entails a more ergonomic posture for the wrist joint. Previous minimally invasive surgical experience is a positive influencing factor on the surgeons' wrist postures during laparoscopy.
机译:目的:本研究旨在通过从数据手套Cyber​​Glove(R)获取信息并确定现有的腕部疾病风险水平,分析使用两种不同的腹腔镜器械手柄期间外科医生的手部空间结构。材料和方法:五十名外科医生参加了这项研究,并分为三组(新手,中级和专家)。每个对象分别在物理模拟器上分别使用轴向处理或环形处理的仪器执行缝合和解剖任务。 Cyber​​Glove(R)记录了手和腕的位置,并采用了改进的RULA方法来确定手腕疾病的适当风险水平。结果:在比较两个任务时,我们发现11个手套式传感器中有7个在统计学上有显着差异。 RULA方法表明,除使用轴向手柄器械的专家外,所有受试者在模拟器上进行缝合和解剖任务时都表现出偏见的手腕姿势。结论:数据手套Cyber​​Glove(R)允许区分使用不同仪器进行的两次腹腔镜锻炼。而且,当使用轴向手柄持针器进行腹腔镜体内缝合时,腕关节的姿势更加符合人体工程学。先前的微创手术经验是腹腔镜检查期间外科医生腕部姿势的积极影响因素。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号