首页> 外文期刊>Microscopy and microanalysis: The official journal of Microscopy Society of America, Microbeam Analysis Society, Microscopical Society of Canada >Development of an Integrated Thermocouple for the Accurate Sample Temperature Measurement During High Temperature Environmental Scanning Electron Microscopy (HT-ESEM) Experiments
【24h】

Development of an Integrated Thermocouple for the Accurate Sample Temperature Measurement During High Temperature Environmental Scanning Electron Microscopy (HT-ESEM) Experiments

机译:开发用于在高温环境扫描电子显微镜(HT-ESEM)实验中精确测量样品温度的集成热电偶

获取原文
获取原文并翻译 | 示例

摘要

We have developed two integrated thermocouple (TC) crucible systems that allow precise measurement of sample temperature when using a furnace associated with an environmental scanning electron microscope (ESEM). Sample temperatures measured with these systems are precise (+/- 5 degrees C) and reliable. The TC crucible systems allow working with solids and liquids (silicate melts or ionic liquids), independent of the gas composition and pressure. These sample holder designs will allow end users to perform experiments at high temperature in the ESEM chamber with high precision control of the sample temperature.