首页> 外文期刊>Microvascular Research: An International Journal >Instrument-independent flux units for laser Doppler perfusion monitoring assessed in a multi-device study on the renal cortex.
【24h】

Instrument-independent flux units for laser Doppler perfusion monitoring assessed in a multi-device study on the renal cortex.

机译:在肾脏皮质的多设备研究中评估了与仪器无关的激光多普勒血流监测通量单位。

获取原文
获取原文并翻译 | 示例
           

摘要

To investigate the feasibility of instrument-independent perfusion units for laser Doppler flowmetry, a comparison was performed of two commercial fiberoptic laser Doppler perfusion monitors measuring the same flux situation for two different types of probes. In vivo measurements were performed on the cortex of pig's kidney, with an ultrasonic arterial flow meter as reference. The flow was mainly varied by internal arterial constriction using a balloon catheter. For each probe, instruments are compared in terms of the ratio of laser Doppler flux and arterial flow. For a given probe, the flux-to-flow ratios of the two instruments show a linear mutual relationship for a wide variety of arterial flows and laser Doppler fluxes. In vitro measurements were performed on an aqueous suspension of polystyrene microspheres. For the probe with interfiber distance 500 microm the ratio of the in vivo fluxes appears to agree within 16% to the value found in vitro, while for the 250-microm probe a difference of 28% was found. For a wide range of fluxes, the in vivo flux values of one instrument can be translated into flux values for the other instrument, in spite of the instrumental differences. This enables the user to render experimental results independent of the specific instrument, thus facilitating multi-center studies.
机译:为了研究用于激光多普勒血流仪的独立于仪器的灌注装置的可行性,对两种商用光纤激光多普勒血流监测仪进行了比较,该监测仪针对两种不同类型的探头测量了相同的通量情况。以超声动脉流量计为参考,在猪肾皮质上进行体内测量。流量主要通过使用球囊导管的内部动脉收缩而变化。对于每个探头,都根据激光多普勒通量和动脉血流的比率对仪器进行比较。对于给定的探头,两种仪器的通量流量比在多种动脉流量和激光多普勒通量之间显示出线性的相互关系。在聚苯乙烯微球的水悬浮液上进行体外测量。对于纤维间距离为500微米的探针,体内通量的比率似乎与在体外发现的值一致,在16%内,而对于250微米探针,发现的差异为28%。对于较大范围的通量,尽管仪器存在差异,但一种仪器的体内通量值可以转换为另一种仪器的通量值。这使用户可以独立于特定仪器呈现实验结果,从而简化了多中心研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号