摘要

This paper presents the design, fabrication and characterization of a novel transient hot-wire device for measurement of thermal conductivity of non-conductive fluids. The key features of the cell are small sample amount, simple fabrication, accurate measurement and transparency, which allows the study of both thermal and optical properties of the fluids. A 20 mu m-diameter platinum hot wire is symmetrically suspended along the central axis of the capsulated cell to avoid the effect from the walls. A four-point resistance measurement method was used to measure the transient resistance with high accuracy. The accuracy of the thermal conductivity measurement was validated with reference values and very good agreement was achieved.
机译:本文介绍了一种用于测量非导电流体导热系数的新型瞬态热线装置的设计,制造和特性。样品池的关键特征是样品量少,制备简单,测量准确且透明,这使得可以研究流体的热学性质和光学性质。一条直径为20μm的铂金热线沿胶囊化电池的中心轴对称悬挂,以避免受到壁的影响。使用四点电阻测量方法来高精度测量瞬态电阻。导热系数测量的准确性已通过参考值进行了验证,并取得了很好的一致性。