首页> 外文期刊>Mikrochimica Acta: An International Journal for Physical and Chemical Methods of Analysis >A novel anti-interference and pH-modulation device: application to enzyme-free glucose detection
【24h】

A novel anti-interference and pH-modulation device: application to enzyme-free glucose detection

机译:一种新型的抗干扰和pH调节装置:在无酶葡萄糖检测中的应用

获取原文
获取原文并翻译 | 示例
           

摘要

We report on a novel anti-interference and pH-modulation device (herein after referred to as "device"). It is based on electrodialysis and can continuously increase the pH value of the carrier solution and - at the same time - remove interfering analytical signals obtained for ascorbic acid (AA) and uric acid (UA). The "device" was coupled to the FIA-amperometric detection of glucose. The linear range is from 1 μmol L~(-1) to 0.4 mmol L~(-1), with a sensitivity of 213 μA cm~(-2) mM~(-1) and a detection limit of 1 μmol L~(-1) at a signal-to-noise ratio of 3. The method was used to sucessfully determine glucose in serum. This study represents a novel technique for overcoming analytical interference and is expected to find applications in liquid chromatography, for example in on-line pH-modulation if different pH values are needed for separation and detection.
机译:我们报告了一种新型的抗干扰和pH调节装置(以下简称“装置”)。它基于电渗析,可不断提高载体溶液的pH值,同时消除对抗坏血酸(AA)和尿酸(UA)的干扰分析信号。 “装置”与FIA-安培法检测葡萄糖结合。线性范围为1μmolL〜(-1)至0.4 mmol L〜(-1),灵敏度为213μAcm〜(-2)mM〜(-1),检测极限为1μmolL〜 (-1)的信噪比为3。该方法用于成功测定血清中的葡萄糖。这项研究代表了一种克服分析干扰的新技术,并且有望在液相色谱中找到应用,例如,如果需要用于分离和检测的不同pH值,则可以用于在线pH调节。