Estimation of Intraocular Pressure in Rabbits Using Noncontact Tonometer: A Comparative Evaluation with Schiotz Tonometer

机译:非接触式眼压计估算家兔眼压:与肖尔兹眼压计的比较评估

获取原文
获取原文并翻译 | 示例
           

摘要

Estimation of intraocular pressure in rabbits is often done by indentation tonometry using Schiotz tonometer. The use of Schiotz tonometer for estimation of intraocular pressure requires the use of topical anesthesia, poses risk of infection and injury to cornea and puts the animal under stress. Therefore, in this study noncontact tonometer was used to estimate intraocular pressure in rabbits and the technique is described for the first time. Twenty-five rabbits were subjected to intraocular pressure estimation using Schiotz as well as noncontact tonometer by two independent observers. All intraobserver and interobserver differences were within two standard deviation from the mean, indicating good repeatability and reproducibility of the two methods. A good linear correlation was observed between measurements done by Schiotz and noncontact tonometers with the correlation coefficient (r~2) of 0.9122 and regression equation y = 1.632x + 4.8684. No significant difference was observed in percent change in intraocular pressure in timolol-treated, water-loaded rabbits using noncontact tonometer and Schiotz tonometer. Noncontact tonometer provides an easy, quick and safe method for intraocular pressure estimation, which is comparable to the conventional method of Schiotz tonometry. Therefore, despite the cost, noncontact tonometry seems to have a greater scope in experimental studies.
机译:通常使用Schiotz眼压计通过压痕眼压计来估计兔眼内压。使用Schiotz眼压计估算眼内压需要使用局部麻醉,存在感染和角膜损伤的风险,并使动物承受压力。因此,在这项研究中,非接触式眼压计用于评估兔眼内压,并且首次描述了该技术。由两名独立的观察者使用Schiotz以及非接触式眼压计对25只兔子进行眼内压评估。观察者间和观察者间的所有差异均在均值的两个标准差以内,表明这两种方法均具有良好的可重复性和可重复性。在Schiotz和非接触式眼压计进行的测量之间观察到良好的线性相关性,相关系数(r〜2)为0.9122,回归方程y = 1.632x + 4.8684。使用非接触式眼压计和Schiotz眼压计,经噻吗洛尔处理的水负荷兔子的眼内压变化百分比没有显着差异。非接触式眼压计提供了一种简便,快速和安全的眼压估计方法,与传统的Schiotz眼压计方法相当。因此,尽管成本高昂,但非接触眼压测量法似乎在实验研究中具有更大的范围。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号