首页> 外文期刊>Microchemical Journal: Devoted to the Application of Microtechniques in all Branches of Science >Determination of bromine by high resolution molecular absorption of strontium mono bromide generated in a graphite furnace
【24h】

Determination of bromine by high resolution molecular absorption of strontium mono bromide generated in a graphite furnace

机译:石墨炉中生成的一溴化锶的高分辨分子吸收法测定溴

获取原文
获取原文并翻译 | 示例

摘要

The bromine concentrationwas determined bymeans ofmolecular absorption of strontiummonobromide (SrBr) generated in the gas phase of the vaporizer of a high-resolution continuumsource electrothermal atomic absorption spectrometer. For this purpose, the wine and drug samples were introduced into the graphite furnace together with Sr and the molecular absorption of SrBr at 651.0581 nm was measured. The effects of SrBr wavelength, graphite furnace program, and the amount of Sr on the accuracy, linearity and sensitivitywere investigated and the experimental parameters were optimized. All quantifications were performed in zirconium coated tubes. The limit of detection and the characteristicmass for themethodwere 1.6 ng and 3.0 ng of Br, respectively. In order to check the risk of non-spectral interference, all quantifications were performed by both standard addition and external calibration techniques. In general, the results were not significantly different. Finally, the brominewas determined in wines and various drug samples obtained frommarkets and drugstores. The bromine quantity in the drug samples was in agreement with those given by the producers.
机译:溴的浓度是由高分辨率连续介质电热原子吸收光谱仪的汽化器气相中生成的一溴化锶(SrBr)的分子吸收率确定的。为此,将葡萄酒和药品样品与Sr一起引入石墨炉,并测量了651.0581 nm处SrBr的分子吸收。研究了SrBr波长,石墨炉程序和Sr含量对精度,线性和灵敏度的影响,并优化了实验参数。所有定量均在锆涂层试管中进行。该方法的检出限和特征质量分别为1.6 ng和3.0 ng Br。为了检查非光谱干扰的风险,所有定量均通过标准添加和外部校准技术进行。通常,结果没有显着差异。最后,在葡萄酒和从市场和药房获得的各种药物样品中测定了溴。药品样品中的溴含量与生产商提供的数量一致。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号