Simple and fast near-infrared spectroscopic analysis of hydroxyl number of polyol using a disposable glass vial

机译:使用一次性玻璃瓶对多元醇的羟值进行简单,快速的近红外光谱分析

获取原文
获取原文并翻译 | 示例
       

摘要

Hydroxyl (OH) number of polyol was measured using near-infrared (NIR) spectroscopy with the use of a disposable glass vial as a sample container. Polyols are viscous, so disposable vials are advantageous when spectroscopic methods are employed. Due to the curvature of the vial walls, a narrow aperture was used to minimize the spectroscopic deviations. The narrow aperture attenuated the NIR radiation and increased the spectral noise in the collected polyol spectra. Wavelet transformation (WT) was employed to reduce this noise and partial least squares (PLS) calibration model was developed. The overall prediction results compare well with those from conventional wet analysis that requires time (1-3 h) and large amounts of chemical reagents. NIR spectroscopy with the use of disposable vials can be utilized for a simple and fast quality assurance of polyol in actual industrial settings. © 2004 Elsevier B.V. All rights reserved.
机译:使用一次性玻璃瓶作为样品容器,使用近红外(NIR)光谱仪测量了多元醇的羟基(OH)值。多元醇是粘性的,因此当使用光谱方法时,一次性小瓶是有利的。由于样品瓶壁的曲率,使用了一个狭窄的孔以使光谱偏差最小。狭窄的孔径使NIR辐射衰减,并增加了收集的多元醇光谱中的光谱噪声。利用小波变换(WT)来减少这种噪声,并开发了偏最小二乘(PLS)校准模型。总体预测结果与需要时间(1-3小时)和大量化学试剂的常规湿法分析结果相比较。使用一次性小瓶的NIR光谱可用于在实际工业环境中简单,快速地保证多元醇的质量。 &复制; 2004 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号