摘要

A transversely heated graphite atomizer (THGA) with longitudinal Zeeman background correction was used for the determination of silver, nickel, cobalt, and chromium in water and human urine samples. The performance of the new type furnace was compared with that of a conventional ''Massmann''-type atomizer. The gas phase and covolatilization interferences of chloride and sulfate matrices in the determination of silver in water observed in the conventional furnace were successfully eliminated using the THGA technique. Detection limits for nickel, cobalt, and chromium were significantly improved using the new atomizer. The results of the determination of trace elements in standard reference materials showed good agreement with the certified or informative values. (C) 1995 Academic Press, Inc. [References: 8]
机译:具有纵向塞曼背景校正的横向加热石墨雾化器(THGA)用于测定水和人尿液样品中的银,镍,钴和铬。将新型熔炉的性能与传统的“ Massmann”型雾化器的性能进行了比较。使用THGA技术成功消除了传统熔炉中观察到的氯化物和硫酸盐基质在测定水中银中的气相和共挥发干扰。使用新的雾化器,镍,钴和铬的检出限大大提高。标准参考物质中痕量元素的测定结果与认证或参考值具有很好的一致性。 (C)1995 Academic Press,Inc. [参考:8]