首页> 外文期刊>Microchemical Journal: Devoted to the Application of Microtechniques in all Branches of Science >Ultraviolet photolysis assisted mineralization and determination of trace levels of Cr, Cd, Cu, Sn, and Pb in isosulfan blue by ICP-MS
【24h】

Ultraviolet photolysis assisted mineralization and determination of trace levels of Cr, Cd, Cu, Sn, and Pb in isosulfan blue by ICP-MS

机译:紫外光解辅助矿化并通过ICP-MS测定异硫蓝中痕量的Cr,Cd,Cu,Sn和Pb

获取原文
获取原文并翻译 | 示例
       

摘要

Isosulfan blue dye is used in sentinel node mapping technique to evaluate breast cancer patients where determination of different trace elements is required. A UV-photolysis assisted mineralization of isosulfan blue is described here for the determination of trace elements (Cr, Cd, Cu, Sn and Pb) by Inductively Coupled Plasma Mass Spectrometry (ICP-MS). In the present study advanced oxidation processes such as UV, UV/H_2O_2, and UV/H_2O_2/HNO_3 have been assessed and compared for the degradation and mineralization of isosulfan blue dye. The extent of mineralization was determined on the basis of total organic carbon (TOC) measurement. The adopted procedure (UV/HNO_3/H_2O_2) provided very low TOC content (0.5%), corresponding to a mineralization efficiency of >99%. With the present procedure, the use of dynamic reaction cell (DRC) or collision cell was not required for chromium determination which suffers interference from the presence of carbon. Method detection limits were 0.0028, 0.0021, 0.016, 0.007, and 0.005μgg~(-1) for Cr, Cd, Cu, Sn and Pb respectively. The values obtained by the proposed method were cross-validated by those obtained by ETAAS analysis. The expanded uncertainties in the measurement at 95% confidence level (coverage factor 2) are in the range of 13.7 -25.8%.
机译:异硫氰酸蓝染料用于前哨淋巴结定位技术,以评估需要确定各种痕量元素的乳腺癌患者。本文介绍了紫外光辅助异硫蓝的矿化方法,用于通过电感耦合等离子体质谱法(ICP-MS)测定痕量元素(Cr,Cd,Cu,Sn和Pb)。在本研究中,已评估并比较了高级氧化过程(例如UV,UV / H_2O_2和UV / H_2O_2 / HNO_3)对异硫蓝染料的降解和矿化作用。矿化程度是根据总有机碳(TOC)测量确定的。所采用的程序(UV / HNO_3 / H_2O_2)的TOC含量非常低(0.5%),对应于> 99%的矿化效率。通过本程序,不需要动态测定池(DRC)或碰撞池进行铬测定,而铬受到碳的干扰。 Cr,Cd,Cu,Sn和Pb的方法检出限分别为0.0028、0.0021、0.016、0.007和0.005μgg〜(-1)。通过提议的方法获得的值与通过ETAAS分析获得的值交叉验证。在95%置信水平(覆盖因子2)下,测量的不确定性扩大到13.7 -25.8%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号