首页> 外文期刊>Microchemical Journal: Devoted to the Application of Microtechniques in all Branches of Science >Determination of Cr(VI) in the presence of Cr(III) and humic acid by cathodic stripping voltammetry
【24h】

Determination of Cr(VI) in the presence of Cr(III) and humic acid by cathodic stripping voltammetry

机译:阴极溶出伏安法测定三价铬和腐植酸存在下的六价铬

获取原文
获取原文并翻译 | 示例
           

摘要

A voltammetric procedure in the flow system for determination of traces of Cr(VI) in the presence of WHO and humic acid is presented. The calibration graph is linear from 5 X 10(-10) to 1 X 10(-7) mol l(-1) for an accumulation time of 120 s. The R.S.D. for 1 X 10(-8) mol l(-1) Cr(VI) is 5.3% (n = 5). The detection limit estimated from 3sigma for a low concentration of Cr(VI) and accumulation time of 120 s is 2 X 10(-10) mol l(-1). The method can be used for Cr(VI) determination in the presence of up to 50 mg l(-1) of humic acid. The validation of the method was carried out by studying the recovery of Cr(VI) from spiked river water and by the comparison of the results of determination of Cr(VI) in a soil sample. The method cannot be used for analysis of samples containing high concentrations of chloride ions such as seawater and estuarine water. (C) 2002 Elsevier Science B.V. All rights reserved. [References: 22]
机译:提出了一种在流动系统中的伏安法,用于在WHO和腐殖酸存在下测定痕量的Cr(VI)。校准曲线从5 X 10(-10)到1 X 10(-7)mol l(-1)是线性的,累积时间为120 s。 R.S.D.对于1 X 10(-8)mol l(-1),Cr(VI)为5.3%(n = 5)。根据3sigma估算,低浓度的Cr(VI)和120s的累积时间的检测限为2 X 10(-10)mol l(-1)。该方法可在高达50 mg l(-1)腐植酸存在下用于Cr(VI)测定。该方法的有效性是通过研究从加标河水中回收的Cr(VI)以及比较土壤样品中Cr(VI)的测定结果来进行的。该方法不能用于分析含有高浓度氯离子的样品,例如海水和河口水。 (C)2002 Elsevier Science B.V.保留所有权利。 [参考:22]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号