Accurate calibration of filter radiometers against a cryogenic radiometer using a trap detector

机译:使用陷阱检测器针对低温辐射计对过滤器辐射计进行准确校准

获取原文
获取原文并翻译 | 示例

摘要

At the Physikalisch-Technische Bundesanstalt filter radiometers are used for the comparison of three primary radiometric standards, an electron storage ring, a black body of known temperature and a cryogenic radiometer. The calibration of the filter radiometers is performed in a two-step process. First, a trap detector is calibrated at the cryogenic radiometer at discrete laser lines. In the second step, this trap detector is used to calibrate the filter radiometers at a spectral comparator. Applying a correction for stray light, the responsivities of the filter radiometers can be measured over a dynamic range of more than eight decades in the wavelength range 420 nm to 1200 nm. To overcome drift effects of the interference filters, the filter radiometers were calibrated every night before and after an application. Following this procedure the responsivity was measured with a relative uncertainty of less than 5×10~(-4) at the 1σ level.
机译:Physikalisch-Technische Bundesanstalt过滤器辐射计用于比较三个主要辐射标准,一个电子存储环,一个已知温度的黑体和一个低温辐射计。滤光片辐射计的校准过程分两步进行。首先,在低温辐射计上以离散激光线对陷阱检测器进行校准。在第二步中,该陷阱检测器用于在光谱比较器处校准滤光片辐射计。通过对杂散光进行校正,可以在波长范围为420 nm至1200 nm的超过八十年的动态范围内测量滤光片辐射计的响应度。为了克服干涉滤光片的漂移效应,在应用前后,每晚都要对滤光片辐射计进行校准。按照此程序,在1σ水平下以相对不确定度小于5×10〜(-4)的方式测量响应度。