【24h】

Measurements using two types of transfer radiometer developed for a monochromator-based cryogenic radiometer facility

机译:使用为基于单色仪的低温辐射计设施开发的两种类型的传输辐射计进行测量

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

A facility for the monochromator-based calibration of transfer standard detectors using a cryogenic radiometer has been set up at the National Research Council of Canada (NRC). A brief status report is given on this facility, which has undergone some modifications since it was first set up in 1994. Measurements so far have concentrated on the spectral range 450 nm to 700 nm. The wavelength calibration of the monochromator used with this facility is discussed in detail. Two types of transfer radiometer have been developed for use with the cryogenic radiometer. The original radiometers use a single photodiode and incorporate a quartz window. The second type 0f radiometer is a windowless three-detector trap using special 18 mm x 18 mm silicon photodiodes. Some of the peoperties of these two types of transfer radiometer are compared: spectral responsivity, uniformity, polarization sensitivity, and linearity. Results of calibrations of these radiometers over one to two years are presented. These measurements suggest that the long-term reproducibility of the calibrations is approximately + 0.025 % for a single radiometer, but close to +0.01% for a group of radiometers. The transfer radiometers can be used to calibrate accurately other detectors using auxiliary apparatus. Measurements are presented which suggest that the transfer uncertainty is close to +0.01% on average. The accuracy of the cryogenic-radiometer-based calibrations can be estimated by linking them to the international comparison of spectral-responsivity scales carried out at the Bureau International des Poids et Mesures in 1994 under the auspices of the Consultative Committee for Photometry and Radiometry (CCPR). From this comparison, it is estimated that the overall accuracy of the cryogenic radiometer calibrations is better than +0.05 % in the range 450 nm to 700 nm.
机译:加拿大国家研究委员会(NRC)已建立了使用低温辐射计对基于单色仪的转移标准检测器进行校准的设施。关于此设施的简要状态报告,自1994年首次建立以来已进行了一些修改。到目前为止,测量集中在450 nm至700 nm的光谱范围内。详细讨论了与此设备一起使用的单色仪的波长校准。已经开发出两种类型的与低温辐射计一起使用的传递辐射计。原始的辐射计使用单个光电二极管并装有石英窗。第二种类型的0f辐射计是使用特殊的18 mm x 18 mm硅光电二极管的无窗三探测器陷阱。比较了这两种类型的传输辐射计的某些功能:光谱响应度,均匀性,极化灵敏度和线性。介绍了这些辐射计在一到两年内的校准结果。这些测量结果表明,对于单个辐射计,校准的长期可重复性约为+ 0.025%,而对于一组辐射计,则接近+ 0.01%。传输辐射计可用于使用辅助设备准确校准其他探测器。提出的测量表明转移不确定性平均接近+ 0.01%。基于低温辐射计的校准的准确性可以通过将其与1994年在国际光度学和辐射度咨询委员会(CCPR)主持下在国际局Poids et Mesures局进行的光谱响应度标度的国际比较相联系来估算。 )。根据该比较,估计在450 nm至700 nm范围内,低温辐射计校准的总体精度优于+0.05%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号