A 400 nm to 2500 nm absolute spectral radiance comparison using filter radiometers

机译:使用滤光辐射计进行400 nm至2500 nm的绝对光谱辐射度比较

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

An integrating-sphere source developed by MTL Systems, Inc. was calibrated for them by a commercial standards laboratory traceable to the National Institute of Standards and Technology (NIST), resulting in values of spectral radiance in the 400 nm to 2500 nm spectral region. The spectral radiance of the MTL sphere source was then measured using three filter radiometers, also NIST-traceable, that were developed for the Earth Observing System (EOS). In the visible and near-infrared, the values from two Si-photodiode filter radiometers and the supplied calibration values agree within their mutual uncertainties, but in the short-wave infrared, the agreement between the filter-radiometer results and the calibration values is less satisfactory. The reason for this discrepancy is not understood, and additional measurements should be performed.
机译:由MTL Systems,Inc.开发的积分球源已通过可追溯至美国国家标准技术研究院(NIST)的商业标准实验室对其进行了校准,从而获得了400 nm至2500 nm光谱范围内的光谱辐射值。然后,使用为地球观测系统(EOS)开发的三台过滤器辐射计(也可溯源至NIST)测量MTL球体源的光谱辐射度。在可见光和近红外中,两个硅光电二极管滤光片辐射计的值和提供的校准值在彼此不确定的范围内一致,但在短波红外中,滤光片辐射计结果与校准值之间的一致性较小满意的。这种差异的原因尚不明确,应进行其他测量。