首页> 外文期刊>Metrologia: International Journal of Scientific Metrology: = Internationale Zeitschrift fur Wissenschaftliche Metrologie: = Journal International de Metrologie Scientifique >Perturbations of the local gravity field due to mass distribution on precise measuring instruments: a numerical method applied to a cold atom gravimeter
【24h】

Perturbations of the local gravity field due to mass distribution on precise measuring instruments: a numerical method applied to a cold atom gravimeter

机译:精确测量仪器上的质量分布引起的局部重力场扰动:应用于冷原子重力仪的数值方法

获取原文
获取原文并翻译 | 示例
           

摘要

We present a numerical method, based on a FEM simulation, for the determination of the gravitational field generated by massive objects, whatever geometry and space mass density they have. The method was applied for the determination of the self-gravity effect of an absolute cold atom gravimeter which aims at a relative uncertainty of 10~(-9). The deduced bias, calculated with a perturbative treatment, is finally presented. The perturbation reaches (1.3 +- 0.1) X 10~(-9) of the Earth's gravitational field.
机译:我们提出了一种基于有限元模拟的数值方法,用于确定由大型物体产生的重力场,无论它们具有什么几何形状和空间质量密度。该方法用于确定绝对冷原子重力仪的自重效应,其相对不确定度为10〜(-9)。最后介绍了用微扰处理计算出的偏差。扰动达到地球引力场的(1.3 +-0.1)X 10〜(-9)。