【24h】

Uncertainty analysis of photometer directional response index f_2 using Monte Carlo simulation

机译:蒙特卡罗模拟法对光度计方向响应指数f_2的不确定度分析

获取原文
获取原文并翻译 | 示例
           

摘要

The uncertainty of the photometer directional response index f_2 is investigated using Monte Carlo simulations and error models derived for the characterization measurements of three photometers. Both random and biased types of errors are used simultaneously to model the measurement noise, drift of the light source, alignment of the photometer, and the errors of the rotary stage. The values of f_2 obtained for each azimuth angle of the photometer are sensitive to the asymmetry of the directional response. It is shown by variation of the photometer symmetry that the changes in the value of f_2 calculated as an average of all simulated values are small. It is further demonstrated that with typical measurement distances the quality index f_2 can be sensitive to the position of the axis of rotation with respect to the receiving plane of the photometer. The simulated f_2 values of the three photometers with expanded uncertainties (k = 2) are (4.325 ± 0.035)%, (1.698 ± 0.027)% and (4.315 ± 0.030)%.
机译:使用蒙特卡洛模拟和为三个光度计的特性测量得出的误差模型,研究了光度计方向响应指数f_2的不确定性。随机误差和偏置误差均同时用于建模测量噪声,光源漂移,光度计对准和旋转平台误差。对于光度计的每个方位角获得的f_2值对方向响应的不对称性敏感。通过光度计对称性的变化表明,作为所有模拟值的平均值计算出的f_2值的变化很小。进一步证明,对于典型的测量距离,质量指标f_2可能对旋转轴相对于光度计接收平面的位置敏感。具有不确定性(k = 2)的三个光度计的模拟f_2值分别为(4.325±0.035)%,(1.698±0.027)%和(4.315±0.030)%。