Laser power-meter comparison at far-infrared wavelengths and terahertz frequencies

机译:远红外波长和太赫兹频率下的激光功率计比较

获取原文
获取原文并翻译 | 示例
           

摘要

We have evaluated the responsivity of seven different thermal detectors compared to an electrically calibrated photoacoustic reference detector at 119 (mu)m (2.5 THz) and 394 (mu)m (0.76 THz) laser wavelengths. Among the thermal detectors is an electrically calibrated thermopile having a vertically aligned carbon nanotube array as the absorber. We document the uncertainty contributions attributable to the photoacoustic reference detector along with a definition of a calibration factor based on the measurement protocol. The expanded relative uncertainty (k velence 2) and a calibration factor of each detector are tabulated.
机译:与电校准的光声参考检测器相比,我们在119μm(2.5 THz)和394μm(0.76 THz)激光波长下评估了七个不同热检测器的响应度。在热探测器中是具有垂直排列的碳纳米管阵列作为吸收体的电校准热电堆。我们记录了归因于光声参考检测器的不确定性贡献以及基于测量协议的校准因子的定义。将每个检测器的扩展相对不确定度(系数2)和校准系数制成表格。