Source and detector calibration in the UV and VUV at BESSY II

机译:在BESSY II的UV和VUV中进行光源和检测器校准

获取原文
获取原文并翻译 | 示例
           

摘要

After the shutdown of the BESSY I electron storage ring, PTB's two normal-incidence monochromator beamlines for the calibration of radiation sources and detectors in the wavelength range from 40 nm to 400 nm were transferred to PTB's new radiometry laboratory at BESSY II. In this paper, the beamlines and their properties are briefly described. First results for the calibration of secondary source standards and detector standards at BESSY II are shown and demonstrate the successful re-establishment of the measurement capabilities. In the ultraviolet, calibration of deuterium lamps was reproduced within 1% and of silicon trap photodetectors within 0.2%.
机译:在BESSY I电子存储环关闭后,PTB的两条用于校准辐射源和检测器(波长范围从40 nm到400 nm)的法线入射单色仪光束线被转移到BESSY II的PTB新的辐射实验室。在本文中,简要描述了束线及其特性。展示了在BESSY II上校准二级源标准品和检测器标准品的初步结果,并证明了测量能力的成功重建。在紫外线下,氘灯的校准值在1%以内,硅阱光电探测器的校准值在0.2%以内。