Procedure for the calibration of surface temperature sensors used in dimensional metrology

机译:尺寸计量中使用的表面温度传感器的校准程序

获取原文
获取原文并翻译 | 示例
           

摘要

A procedure for the calibration of sensors used for the measurement ofsurface temperature in dimensional metrology is developed. It consists of the utilization of a climate chamber, with an aluminium block inside, to house reference and calibrating sensors. The temperature of the sensors is controlled by the climate chamber and uncertainties are reduced due to the utilization of the aluminium block. The method is demonstrated by comparing the results obtained with those of a water bath calibration.
机译:开发了用于在尺寸计量中测量表面温度的传感器的校准程序。它包括利用气候室和内部的铝块来容纳参考和校准传感器。传感器的温度由气候室控制,由于使用了铝块,减少了不确定性。通过将获得的结果与水浴校准的结果进行比较,证明了该方法。