Systematic review of uncalibrated arterial pressure waveform analysis to determine cardiac output and stroke volume variation

机译:系统检查未校准的动脉压波形分析以确定心输出量和中风量变化

获取原文
获取原文并翻译 | 示例
       

摘要

SummaryThe FloTrac/Vigileo?, introduced in 2005, uses arterial pressure waveform analysis to calculate cardiac output (CO) and stroke volume variation (SVV) without external calibration. The aim of this systematic review is to evaluate the performance of the system. Sixty-five full manuscripts on validation of CO measurements in humans, published in English, were retrieved; these included 2234 patients and 44 592 observations. Results have been analysed according to underlying patient conditions, that is, general critical illness and surgery as normodynamic conditions, cardiac and (post)cardiac surgery as hypodynamic conditions, and liver surgery and sepsis as hyperdynamic conditions, and subsequently released software versions. Eight studies compared SVV with other dynamic indices. CO, bias, precision, %error, correlation, and concordance differed among underlying conditions, subsequent software versions, and their interactions, suggesting increasing accuracy and precision, particularly in hypo- and normodynamic conditions. The bias and the trending capacity remain dependent on (changes in) vascular tone with most recent software. The SVV only moderately agreed with other dynamic indices, although it was helpful in predicting fluid responsiveness in 85% of studies addressing this. Since its introduction, the performance of uncalibrated FloTrac/Vigileo? has improved particularly in hypo- and normodynamic conditions. A %error at or below 30% with most recent software allows sufficiently accurate and precise CO measurements and trending for routine clinical use in normo- and hypodynamic conditions, in the absence of large changes in vascular tone.
机译:总结2005年推出的FloTrac / Vigileo?使用动脉压力波形分析无需外部校准即可计算心输出量(CO)和每搏量变化(SVV)。此系统审查的目的是评估系统的性能。检索了以英文出版的关于验证人体中CO含量的65份完整手稿;其中包括2234例患者和44592例观察结果。根据潜在的患者状况(即一般危重疾病和手术为正常动力状况,心脏和(心脏)手术为低动力状况,肝脏手术和败血症为高动力状况)对结果进行了分析,并随后发布了软件版本。八项研究将SVV与其他动态指标进行了比较。 CO,偏差,精度,误差%,相关性和一致性在基础条件,后续软件版本及其相互作用之间有所不同,这表明准确性和精度不断提高,尤其是在低动力和正常动力条件下。偏差和趋势容量仍取决于最新软件的血管张力(变化)。 SVV仅与其他动态指标适度吻合,尽管它有助于预测针对这一问题的研究中有85%的流体反应性。自推出以来,未经校准的FloTrac / Vigileo的性能如何?尤其在血流动力学和正常动力学条件下有所改善。使用最新的软件,如果误差在30%或以下,则可以在正常和动力不足的情况下进行足够准确准确的一氧化碳测量并趋向于常规临床应用,而无需大幅度改变血管张力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号