Quality Control and Validation of Oscillometric Blood Pressure Measurements Taken During an Epidemiological Investigation

机译:在流行病学调查期间进行示波血压测量的质量控制和验证

获取原文
获取原文并翻译 | 示例
           

摘要

This study aims to validate blood pressure (BP) values measured by an oscillometric BP monitor and seek possible calibration methods if discrepancies exist.Noninvasive BP measurement outcomes were determined using an oscillometric BP monitor (Omron HBP-1300) versus a mercury sphygmomanometer (standard device). Two percent of subjects enrolled in an epidemiological investigation were systematically sampled in this study. Intraclass correlation coefficient (ICC) was used to evaluate measurement reliability, paired t-test was used to evaluate trueness, and linear regression was used for calibration. The Association for the Advancement of Medical Instrumentation (AAMI) standards and British Hypertension Society (BHS) protocols were used for validation quality assessment.Both mercury sphygmomanometer (standard device) and oscillometric BP monitor (test device) displayed high reliability. A significant difference in systolic blood pressure (SBP) was observed between devices. SBP calibration was achieved by using an effective linear regression model (B=0.803 and constant=19.592, P<0.001). The calibrated model was corroborated by verification samples (P=0.120) and was found to pass AAMI standards and BHS protocol requirements.Calibrated SBP measurements from the Omron HBP-1300 device were valid. Use of a combination of statistical methods, such as ICC for reliability assessment as well as paired t-test for trueness evaluation can be used to validate data from the oscillometric BP monitors.
机译:这项研究旨在验证示波BP监测仪测量的血压(BP)值并寻找可能的校准方法(如果存在差异)。使用示波BP监测仪(Omron HBP-1300)与水银血压计(标准设备)来确定无创BP测量结果)。在这项研究中,系统地抽取了2%参加流行病学调查的受试者。类内相关系数(ICC)用于评估测量的可靠性,配对t检验用于评估真实性,而线性回归用于校准。使用医疗器械先进协会(AAMI)标准和英国高血压协会(BHS)协议进行质量验证评估,汞血压计(标准设备)和示波BP监测仪(测试设备)均显示出高可靠性。装置之间观察到收缩压(SBP)的显着差异。通过使用有效的线性回归模型(B = 0.803和常数= 19.592,P <0.001)实现SBP校准。校准后的模型得到了验证样品的证实(P = 0.120),并通过了AAMI标准和BHS协议要求。OmronHBP-1300设备的校准SBP测量值是有效的。可以使用诸如ICC之类的统计方法组合进行可靠性评估,以及将t配对检验用于真实性评估,这些方法可以用于验证示波BP监护仪的数据。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号