首页> 外文期刊>Medical engineering & physics. >Fluorescent strip-lights as a source of error in tissue reflectance spectroscopy.
【24h】

Fluorescent strip-lights as a source of error in tissue reflectance spectroscopy.

机译:荧光灯是组织反射光谱中误差的来源。

获取原文
获取原文并翻译 | 示例

摘要

Tissue reflectance spectroscopy is a widely used technique for measuring the blood content of the skin. Fluorescent strip-lighting of a type in widespread use (Osram L36 W/23 Gelbweiss-white) is a potential source of misleading artefacts in tissue reflectance spectroscopy. The fluorescent light signal has successive peaks and troughs in the 520-580 nm range which closely emulate the inverse of the reflected signal of white light modulated by blood. The presence of fluorescent light in the reflected signal cancels some of the blood signature. Conversely, the presence of fluorescent light in the white reference augments the true blood reflected signal. The marked similarity of the resultant artefact blood spectrum to the true blood spectrum is a hazard, as the presence of the artefact may go undetected.
机译:组织反射光谱法是一种广泛用于测量皮肤血​​液含量的技术。广泛使用的一种荧光灯带照明(欧司朗L36 W / 23 Gelbweiss-white)是在组织反射光谱中产生误导性伪像的潜在来源。荧光信号在520-580 nm范围内具有连续的波峰和波谷,这些波峰和波谷与血液调制的白光反射信号的逆过程非常相似。反射信号中荧光的存在会抵消一些血液特征。相反,白色参考中荧光灯的存在会增加真实的血液反射信号。由于伪造物的存在可能未被发现,因此所获得的伪造物光谱与真实血液光谱的显着相似性是一种危害。