首页> 外文期刊>Medical engineering & physics. >Is the 'instant flow' option on the NIOX analyser needed?
【24h】

Is the 'instant flow' option on the NIOX analyser needed?

机译:NIOX分析仪上是否需要“即时流量”选项?

获取原文
获取原文并翻译 | 示例
       

摘要

Asthmatics have increased gaseous nitric oxide in their exhaled breath. Measurements of the fraction of nitric oxide in exhaled breath (FE(NO)) are used for the diagnosis and management of asthma. One analyser (the NIOX chemiluminescence analyser) has a feature (restriction of instant flow, IF) which we suspected was responsible for reducing the success rate for obtaining FE(NO) measurements whilst not affecting the measurement itself. In a study involving 38 children, median age 10 years, we observed that FE(NO) values taken in the same individual with and without IF restriction did not differ. We also observed that the success in obtaining a FE(NO) measurement was higher (57%) when IF was not restricted and was lower (41%) when IF was restricted, p=0.002. Our findings do not support the use of restriction of IF on the NIOX analyser.
机译:哮喘患者呼出气中一氧化氮的含量增加。呼出气中一氧化氮含量的测量(FE(NO))用于哮喘的诊断和治疗。一种分析仪(NIOX化学发光分析仪)具有一项功能(即时流量限制,IF),我们怀疑该功能可降低获得FE(NO)测量值的成功率,同时又不影响测量值本身。在一项涉及38名中位年龄为10岁的儿童的研究中,我们观察到在有和没有IF限制的情况下,同一人的FE(NO)值没有差异。我们还观察到,如果不限制IF,获得FE(NO)测量值的成功率较高(57%),而如果IF受限制,则成功率较低(41%),p = 0.002。我们的发现不支持在NIOX分析仪上使用IF限制。