首页> 外文期刊>Medical Physics >Electrostatic focal spot correction for x-ray tubes operating in strong magnetic fields
【24h】

Electrostatic focal spot correction for x-ray tubes operating in strong magnetic fields

机译:在强磁场下工作的X射线管的静电焦点校正

获取原文
获取原文并翻译 | 示例

摘要

Purpose: A close proximity hybrid x-ray/magnetic resonance (XMR) imaging system offers several critical advantages over current XMR system installations that have large separation distances (~5 m) between the imaging fields of view. The two imaging systems can be placed in close proximity to each other if an x-ray tube can be designed to be immune to the magnetic fringe fields outside of the MR bore. One of the major obstacles to robust x-ray tube design is correcting for the effects of the MR fringe field on the x-ray tube focal spot. Any fringe field component orthogonal to the x-ray tube electric field leads to electron drift altering the path of the electron trajectories.
机译:目的:与当前XMR系统安装相比,近距离混合X射线/磁共振(XMR)成像系统具有几个关键优势,而当前XMR系统的安装在成像视场之间具有较大的分隔距离(〜5 m)。如果可以将X射线管设计为不受MR孔外部的磁条纹场的影响,则可以将两个成像系统彼此靠近放置。坚固的X射线管设计的主要障碍之一是要校正MR​​边缘场对X射线管焦点的影响。垂直于X射线管电场的任何边缘场分量都会导致电子漂移,从而改变电子轨迹的路径。