Performance characterization of a MVCT scanner using multislice thick, segmented cadmium tungstate-photodiode detectors.

机译:使用多层厚的分段钨酸镉光电二极管检测器进行MVCT扫描仪的性能表征。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: Megavoltage computed tomography (MVCT) and megavoltage cone beam computed tomography (MVCBCT) can be used for visualizing anatomical structures prior to radiation therapy treatments to assist in patient setup and target localization. These systems are less susceptible to metal artifacts and provide better CT number linearity than conventional CT scanners. However, their contrast is limited by the properties of the megavoltage photons and the low detective quantum efficiency (DQE) of flat panel detector systems currently available. By using higher DQE, thick, segmented cadmium tungstate detectors, the authors can improve the low contrast detectability of a MVCT system. This in turn would permit greater soft tissue visualization for a given radiation dose, allowing MVCT to be used in more clinical situations. METHODS: This article describes the evaluation of our prototype system that uses thick, segmented detectors. In order to create images using a dose that would be acceptable for day to day patient imaging, the authors evaluated their system using the low intensity bremsstrahlung component of a 6 MeV electron beam. The system was evaluated for its uniformity, high contrast resolution, low contrast detectability, signal to noise ratio, contrast to noise ratio, and CT number linearity. RESULTS: The prototype system was found to have a high contrast spatial resolution of about 5 line pairs per cm, and to be able to visualize a 15 mm 1.5% contrast target with 2 cGy of radiation dose delivered. SNR2 vs radiation dose and mean pixel value vs electron density curves were linear. CONCLUSIONS: This prototype system shows a large improvement in low contrast detectability over current MVCBCT systems.
机译:目的:兆伏计算机断层扫描(MVCT)和兆伏锥束计算机断层扫描(MVCBCT)可用于在放射治疗之前可视化解剖结构,以帮助患者设置和目标定位。与传统的CT扫描仪相比,这些系统不易受到金属伪影的影响,并提供更好的CT编号线性度。但是,它们的对比度受到兆伏光子的特性和当前可用的平板探测器系统的低探测量子效率(DQE)的限制。通过使用更高的DQE,厚的分段钨酸镉检测器,作者可以提高MVCT系统的低对比度可检测性。反过来,对于给定的辐射剂量,这将允许更大的软组织可视化,从而允许将MVCT用于更多的临床情况。方法:本文介绍了对使用厚壁,分段检测器的原型系统的评估。为了使用对患者的日常成像可接受的剂量来创建图像,作者使用6 MeV电子束的低强度component致辐射分量对其系统进行了评估。对系统的均匀性,高对比度分辨率,低对比度可检测性,信噪比,对比度与噪声比以及CT数线性进行了评估。结果:发现原型系统具有约每厘米5条线对的高对比度空间分辨率,并且能够以2 cGy的放射剂量可视化15毫米1.5%对比度目标。 SNR2对辐射剂量以及平均像素值对电子密度曲线是线性的。结论:该原型系统在低对比度可检测性方面比当前的MVCBCT系统有很大的提高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号