首页> 外文期刊>Medical Physics >Experimental investigation of dose calibrator response for 125I brachytherapy solutions contained in 5 mL plastic syringes and 2 mL conical glass v-vials as a function of filling mass.
【24h】

Experimental investigation of dose calibrator response for 125I brachytherapy solutions contained in 5 mL plastic syringes and 2 mL conical glass v-vials as a function of filling mass.

机译:作为填充量的函数,对5 mL塑料注射器和2 mL锥形玻璃v型瓶中所含125 I近距离治疗溶液的剂量校准物响应的实验研究。

获取原文
获取原文并翻译 | 示例
           

摘要

The effect of measurement geometry on the determination of the activity of solutions containing 125I for use in brachytherapy applications has been investigated for 5 mL plastic syringes and 2 mL conical glass dose vials as a function of filling mass. New dial settings for the syringes over a filling mass range of 1 to 3 g have been determined to be 497+/-8 and 469+/-8 (expanded, k = 2, uncertainties) for the NIST Capintec CRC- 12 and Capintec CRC-35R, respectively, with any effect due to the filling mass lying within the uncertainty in the activity calibration. A filling mass effect was observed in the dose vials, causing a 10.5% reduction in the chamber response from a 2 g filling mass to 1 g. Dial settings at 2 g were experimentally found to be 143+/-2 and 135+/-2 (expanded uncertainties) for the NIST Capintec CRC-12 and Capintec CRC-35R, respectively. The appropriate dial settings for the same vials with a 1 g filling mass were found to be 120+/-2 and 114+/-2 for CRC-12 and CRC-35R, respectively. Differences of up to +/-45% in the activity determination were observed between values obtained with the manufacturer's recommended setting and the settings obtained experimentally for each specific geometry. Calibration factors were also determined for a Vinten 671 Radionuclide Calibrator, giving values of 0.226+/-0.009 pA x MBq(-1) and 0.231+/-0.004 pA x MBq(-1) (expanded uncertainties), respectively, for the 1 and 2 g dispensings. This study demonstrates that experimentally determined calibration factors for the exact measurement geometry are necessary when measuring radionuclides in configurations other than the manufacturer's standard geometry, especially when nuclides that emit low-energy radiations are involved.
机译:对于5 mL塑料注射器和2 mL锥形玻璃剂量小瓶,根据填充质量,研究了测量几何形状对用于近距离放射治疗的含125 I溶液活性测定的影响。对于NIST Capintec CRC-12和Capintec,在1到3 g填充质量范围内的注射器的新刻度盘设置已确定为497 +/- 8和469 +/- 8(扩展,k = 2,不确定性)。 CRC-35R分别由于填充量在活性校准不确定性范围内而具有任何影响。在剂量瓶中观察到填充质量效应,使腔室响应从2 g填充质量降低到1 g,降低了10.5%。实验发现,NIST Capintec CRC-12和Capintec CRC-35R在2 g时的刻度盘设置分别为143 +/- 2和135 +/- 2(不确定性扩大)。对于填充质量为1 g的相同样品瓶,适当的刻度盘设置对于CRC-12和CRC-35R分别为120 +/- 2和114 +/- 2。在使用制造商的推荐设置获得的值与通过实验为每种特定几何形状获得的设置之间,观察到活性测定中的差异高达+/- 45%。还针对Vinten 671放射性核素校准器确定了校准系数,对于1校准器,分别给出了0.226 +/- 0.009 pA x MBq(-1)和0.231 +/- 0.004 pA x MBq(-1)的值(不确定性扩大)和2克配药。这项研究表明,当以制造商的标准几何形状以外的其他形式测量放射性核素时,尤其是当涉及发射低能辐射的核素时,必须通过实验确定准确的测量几何形状的校准因子。

著录项

相似文献

  • 外文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号