Purpose: An original radionuclide calibrator method for activity determination is presented. The method could be used for intercomparison surveys for short half-life radioactive sources used in Nuclear Medicine, such as ~(99m)Tc or most positron emission tomography radiopharmaceuticals. Methods: By evaluation of the resulting net optical density (netOD) using a standardized scanning method of irradiated Gafchromic XRQA2 film, a comparison of the netOD measurement with a previously determined calibration curve can be made and the difference between the tested radionuclide calibrator and a radionuclide calibrator used as reference device can be calculated. To estimate the total expected measurement uncertainties, a careful analysis of the methodology, for the case of ~(99m)Tc, was performed: reproducibility determination, scanning conditions, and possible fadeout effects. Since every factor of the activity measurement procedure can influence the final result, the method also evaluates correct syringe positioning inside the radionuclide calibrator.Results: As an alternative to using a calibrated source sent to the surveyed site, which requires a relatively long half-life of the nuclide, or sending a portable calibrated radionuclide calibrator, the proposed method uses a source prepared in situ. An indirect activity determination is achieved by the irradiation of a radiochromic film using ~(99m)Tc under strictly controlled conditions, and cumulated activity calculation from the initial activity and total irradiation time. The irradiated Gafchromic film and the irradiator, without the source, can then be sent to a National Metrology Institute for evaluation of the results.Conclusions: The methodology described in this paper showed to have a good potential for accurate (3%) radionuclide calibrators intercomparison studies for ~(99m)Tc between Nuclear Medicine centers without source transfer and can easily be adapted to other short half-life radionuclides.
机译:目的:提出了一种用于放射性测定的原始放射性核素校准方法。该方法可用于核医学中使用的短半衰期放射源(例如〜(99m)Tc或大多数正电子发射断层扫描放射药物)的比对调查。方法:通过使用辐照的Gafchromic XRQA2膜的标准化扫描方法评估所得的净光密度(netOD),可以将netOD测量值与先前确定的校准曲线进行比较,以及被测放射性核素校准物和放射性核素之间的差异可以计算用作参考设备的校准器。为了估计总的预期测量不确定度,对〜(99m)Tc的方法进行了仔细的分析:重现性确定,扫描条件和可能的淡出效果。由于活动性测量程序的每个因素都会影响最终结果,因此该方法还可以评估注射器在放射性核素校准器中的正确位置。结果:作为替代使用已发送到被调查地点的校准源的方法,这需要较长的半衰期对于核素,或发送便携式校准的放射性核素校正剂,建议的方法使用就地制备的放射源。通过在严格控制的条件下使用〜(99m)Tc辐照放射性致变色膜,并根据初始活性和总辐照时间计算累积活性,可以确定间接活性。然后,可以将辐照的全色致变色膜和辐照器(不带光源)发送给国家计量研究所,以评估结果。结论:本文描述的方法显示出进行精确(3%)放射性核素校准剂比对的良好潜力。无需转移源即可在核医学中心之间进行〜(99m)Tc的研究,并且可以轻松地适用于其他短半衰期放射性核素。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号