首页> 外文期刊>Medical Physics >Comparison of an amorphous silicon/cesium iodide flat-panel digital chest radiography system with screen/film and computed radiography systems--a contrast-detail phantom study.
【24h】

Comparison of an amorphous silicon/cesium iodide flat-panel digital chest radiography system with screen/film and computed radiography systems--a contrast-detail phantom study.

机译:带有屏幕/胶片和计算机射线照相系统的非晶硅/碘化铯平板数字胸片照相系统的比较-对比-细节体模研究。

获取原文
获取原文并翻译 | 示例
       

摘要

Flat-panel (FP) based digital radiography systems have recently been introduced as a new and improved digital radiography technology; it is important to evaluate and compare this new technology with currently widely used conventional screen/film (SF) and computed radiography (CR) techniques. In this study, the low-contrast performance of an amorphous silicon/cesium iodide (aSi/Csl)-based flat-panel digital chest radiography system is compared to those of a screen/film and a computed radiography system by measuring their contrast-detail curves. Also studied were the effects of image enhancement in printing the digital images and dependence on kVp and incident exposure. It was found that the FP system demonstrated significantly better low-contrast performance than the SF or CR systems. It was estimated that a dose savings of 70%-90% could be achieved to match the low-contrast performance of the FP images to that of the SF images. This dose saving was also found to increase with the object size. No significant difference was observed in low-contrast performances between the SF and CR systems. The use of clinical enhancement protocols for printing digital images was found to be essential and result in better low-contrast performance. No significant effects were observed for different kVps. From the results of this contrast-detail phantom study, the aSi/CsI-based flat-panel digital chest system should perform better under clinical situations for detection of low-contrast objects such as lung nodules. However, proper processing prior to printing would be essential to realizing this better performance.
机译:最近,基于平板(FP)的数字X射线照相系统已被引入,作为一种新的改进的数字X射线照相技术。评估此新技术并将其与当前广泛使用的常规屏幕/胶片(SF)和计算机射线照相(CR)技术进行比较非常重要。在这项研究中,通过测量对比度细节,将基于非晶硅/碘化铯(aSi / Csl)的平板数字胸部放射线照相系统与屏幕/胶片和计算机放射线照相系统的低对比度性能进行了比较。曲线。还研究了图像增强在打印数字图像中的效果以及对kVp和入射曝光的依赖性。发现FP系统显示出比SF或CR系统明显更好的低对比度性能。据估计,可以实现70%-90%的剂量节省,以使FP图像的低对比度性能与SF图像的低对比度性能相匹配。还发现这种节省剂量随着物体尺寸的增加而增加。 SF和CR系统之间在低对比度性能方面未观察到显着差异。已发现使用临床增强协议来打印数字图像非常重要,并且可以产生更好的低对比度性能。对于不同的kVps,没有观察到明显的影响。从该对比细节模型研究的结果来看,基于aSi / CsI的平板数字胸部系统在临床情况下应能更好地检测低对比度的物体,例如肺结节。但是,打印之前进行适当的处​​理对于实现此更好的性能至关重要。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号