首页> 外文期刊>Medical Physics >Simulation of dose reduction in tomosynthesis.
【24h】

Simulation of dose reduction in tomosynthesis.

机译:层析合成中剂量减少的模拟。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: Methods for simulating dose reduction are valuable tools in the work of optimizing radiographic examinations. Using such methods, clinical images can be simulated to have been collected at other, lower, dose levels without the need of additional patient exposure. A recent technology introduced to healthcare that needs optimization is tomosynthesis, where a number of low-dose projection images collected at different angles is used to reconstruct section images of an imaged object. The aim of the present work was to develop a method of simulating dose reduction for digital radiographic systems, suitable for tomosynthesis. METHODS: The developed method uses information about the noise power spectrum (NPS) at the original dose level and the simulated dose level to create a noise image that is added to the original image to produce an image that has the same noise properties as an image actually collected at the simulated dose level. As the detective quantum efficiency (DQE) of digital detectors operating at the low dose levels used for tomosynthesis may show a strong dependency on the dose level, it is important that a method for simulating dose reduction for tomosynthesis takes this dependency into account. By applying an experimentally determined relationship between pixel mean and pixel variance, variations in both dose and DQE in relevant dose ranges are taken into account. RESULTS: The developed method was tested on a chest tomosynthesis system and was shown to produce NPS of simulated dose-reduced projection images that agreed well with the NPS of images actually collected at the simulated dose level. The simulated dose reduction method was also applied to tomosynthesis examinations of an anthropomorphic chest phantom, and the obtained noise in the reconstructed section images was very similar to that of an examination actually performed at the simulated dose level. CONCLUSIONS: In conclusion, the present article describes a method for simulating dose reduction suitable for tomosynthesis. However, the method applies equally well to any digital radiographic system, although the benefits of correcting for DQE variations may be smaller.
机译:目的:模拟剂量减少的方法是优化射线照相检查工作中的宝贵工具。使用这种方法,可以模拟临床图像以其他较低剂量水平收集,而无需额外的患者暴露。引入到医疗保健中需要进行优化的最新技术是断层合成,其中以不同角度收集的许多低剂量投影图像用于重建成像对象的断面图像。本工作的目的是开发一种适用于断层合成的数字射线照相系统模拟剂量减少的方法。方法:所开发的方法使用有关原始剂量水平和模拟剂量水平的噪声功率谱(NPS)的信息来创建噪声图像,该噪声图像被添加到原始图像中以产生具有与图像相同的噪声特性的图像实际以模拟剂量水平收集。由于在用于层析合成的低剂量水平下运行的数字检测器的探测量子效率(DQE)可能显示出对剂量水平的强烈依赖性,因此,重要的是,模拟用于层析合成的剂量减少的方法必须考虑到这种依赖性。通过在像素均值和像素方差之间应用实验确定的关系,可以考虑剂量和DQE在相关剂量范围内的变化。结果:该方法在胸部断层合成系统上进行了测试,结果显示可产生模拟剂量降低的投影图像的NPS,该图像与以模拟剂量水平实际采集的图像的NPS十分吻合。模拟剂量减少方法还应用于拟人化的胸部幻影的断层合成检查,并且在重建断面图像中获得的噪声与在模拟剂量水平下实际进行的检查非常相似。结论:总而言之,本文描述了一种适合于断层合成的模拟剂量减少的方法。但是,该方法同样适用于任何数字射线照相系统,尽管校正DQE变化的好处可能较小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号