Magnetic resonance imaging of thermal coagulation effects in a phantom for calibrating thermal therapy devices.

机译:用于校准热疗设备的幻像中的热凝结效应的磁共振成像。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A material has been developed and tested that permanently records thermal response patterns from heating devices. The material consists of a mixture of polyacrylamide and 18% w/w bovine serum albumin. Thermal denaturation is complete when the local temperature exceeds 70 degrees C, causing a large reduction in the T2 of the material. Three-dimensional distributions of "thermal damage" can be assessed using standard magnetic resonance imaging sequences. The material works well with microwave heating devices and is adaptable for use with ultrasound, radio-frequency, or laser heating devices. Suggested uses include characterizing heating devices prior to treatment and developing new clinical applications for thermal therapies.
机译:已经开发并测试了一种材料,该材料可以永久记录加热设备的热响应模式。该材料由聚丙烯酰胺和18%w / w牛血清白蛋白的混合物组成。当局部温度超过70摄氏度时,热变性完成,从而导致材料的T2大大降低。可以使用标准的磁共振成像序列来评估“热损伤”的三维分布。该材料可与微波加热设备一起很好地工作,并适合与超声,射频或激光加热设备一起使用。建议的用途包括在治疗前表征加热装置,并开发热疗法的新临床应用。