Three types of films, Kodak EDR2, Gafchromic EBT, and Gafchromic MD-V2-55, were used to measure relative output factors of 4 and 8 mm collimators of the Leksell Gamma Knife Perfexion. The optical density to dose calibration curve for each of the film types was obtained by exposing the films to a range of known doses. Ten data points were acquired for each of the calibration curves in the dose ranges from 0 to 4 Gy, 0 to 8 Gy, and 0 to 80 Gy for Kodak EDR2, Gafchromic EBT, and Gafchromic MD-V2-55 films, respectively. For the measurement of relative output factors, five films of each film type were exposed to a known dose. All films were scanned using EPSON EXPRESSION 10000 XL scanner with 200 dpi resolution in 16 bit gray scale for EDR2 film and 48 bit color scale for Gafchromic films. The scanned images were imported in the red channel for both Gafchromic films. The background corrections from an unexposed film were applied to all films. The output factors obtained from film measurements were in a close agreement both with the Monte Carlo calculated values of 0.924 and 0.805 for 8 and 4 mm collimators, respectively. These values are provided by the vendor and used as default values in the vendor's treatment planning system. The largest differences were noted for the Kodak EDR 2 films (-2.1% and -4.5% for 8 and 4 mm collimators, respectively). The best agreement observed was for EBT Gafchromic film (-0.8% and +0.6% differences for 8 and 4 mm collimators, respectively). Based on the present values, no changes in the default relative output factor values were made in the treatment planning system.
机译:三种类型的胶片,柯达EDR2,Gafchromic EBT和Gafchromic MD-V2-55,用于测量Leksell伽玛刀Perfexion的4和8 mm准直器的相对输出因子。通过将膜暴露于一系列已知剂量下,可获得每种膜类型的光密度-剂量校准曲线。对于柯达EDR2,Gafchromic EBT和Gafchromic MD-V2-55胶片,剂量曲线分别从0到4 Gy,0到8 Gy和0到80 Gy采集了10个数据点。为了测量相对输出因子,将每种膜类型的五张膜暴露于已知剂量。使用EPSON EXPRESSION 10000 XL扫描仪对所有胶片进行扫描,分辨率为200 dpi,对于EDR2胶片为16位灰度,对于Gafchromic胶片为48位色度。扫描的图像都导入了红色通道中的两种全色胶片。将未曝光胶片的背景校正应用于所有胶片。从胶片测量获得的输出因子与8毫米和4毫米准直器的蒙特卡洛计算值分别为0.924和0.805密切相关。这些值由供应商提供,并用作供应商的治疗计划系统中的默认值。柯达EDR 2胶片的差异最大(8毫米和4毫米准直仪分别为-2.1%和-4.5%)。观察到的最佳协议是EBT全色膜(8和4 mm准直器的差异分别为-0.8%和+ 0.6%)。基于当前值,在治疗计划系统中未更改默认相对输出因子值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号