Multispectral breast imaging using a ten-wavelength, 64 x 64 source/detector channels silicon photodiode-based diffuse optical tomography system.

机译:使用十波长,64 x 64源/检测器通道的多光谱乳房成像,基于硅光电二极管的漫射光学层析成像系统。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

We describe a compact diffuse optical tomography system specifically designed for breast imaging. The system consists of 64 silicon photodiode detectors, 64 excitation points, and 10 diode lasers in the near-infrared region, allowing multispectral, three-dimensional optical imaging of breast tissue. We also detail the system performance and optimization through a calibration procedure. The system is evaluated using tissue-like phantom experiments and an in vivo clinic experiment. Quantitative two-dimensional (2D) and three-dimensional (3D) images of absorption and reduced scattering coefficients are obtained from these experiments. The ten-wavelength spectra of the extracted reduced scattering coefficient enable quantitative morphological images to be reconstructed with this system. From the in vivo clinic experiment, functional images including deoxyhemoglobin, oxyhemoglobin, and water concentration are recovered and tumors are detected with correct size and position compared with the mammography.
机译:我们描述了一种专为乳房成像而设计的紧凑型漫射光学层析成像系统。该系统由64个硅光电二极管检测器,64个激发点和10个在近红外区域的二极管激光器组成,可对乳房组织进行多光谱,三维光学成像。我们还将通过校准程序详细介绍系统性能和优化。该系统使用组织样体模实验和体内临床实验进行评估。从这些实验中获得了吸收性和散射系数降低的定量二维(2D)和三维(3D)图像。提取的降低的散射系数的十个波长光谱使得可以使用该系统重建定量形态图像。通过体内临床实验,可以恢复包括脱氧血红蛋白,氧合血红蛋白和水浓度在内的功能图像,并且与乳房X线摄影术相比,可以以正确的大小和位置检测到肿瘤。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号