Determination of penumbral widths from ion chamber measurements.

机译:通过离子室测量确定半影宽度。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

We have investigated methods of reconstructing beam profiles in the penumbral region using a set of axially symmetric chambers, differing only in the detector radius. In principal, the transfer functions, or kernels, of such chambers should be functions of radius only. Three chambers of radii 0.297, 0.556, and 0.714 cm have been used. The transfer functions of the chambers can be determined by deconvolving the profiles measured with each detector with the PPMC profile. The results indicate that the transfer functions can be parametrized accurately as a Gaussian cutoff at 1.75(r), with (r) the average radius of the chamber. Deconvolution of the measured profiles with the transfer functions yields a profile that agrees with the PPMC profile to +/- 0.5 mm over the 20-80% penumbra.
机译:我们研究了使用一组轴向对称的腔室重建半影区域内光束轮廓的方法,这些腔室仅在探测器半径上有所不同。原则上,这种腔室的传递函数或核应仅是半径的函数。已使用三个半径为0.297、0.556和0.714厘米的腔室。可以通过将每个检测器测得的轮廓与PPMC轮廓解卷积来确定腔室的传递函数。结果表明,传递函数可以精确地参数化为高斯截止点(1.75(r),其中(r)室的平均半径)。用传递函数对测量轮廓进行反卷积得到的轮廓与PPMC轮廓一致,在20-80%的半影上为+/- 0.5 mm。