Dose enhancement close to platinum implants for the 4, 6, and 10 MV stereotactic radiosurgery.

机译:4 MV,6 MV和10 MV立体定向放射外科手术的剂量增加接近铂植入物。

获取原文
获取原文并翻译 | 示例
           

摘要

Three photon interaction processes, namely, the photoelectric effect, Compton effect, and pair production, can occur when materials with high atomic numbers are irradiated by the high- and low-energy bremsstrahlung photons from a linear accelerator. A dose enhancement, due to the photoelectric effect and pair production, near targets with platinum implants (with a high atomic number) in radiosurgery cannot be predicted by the XKnife radiosurgery treatment planning system. In the present work, Monte Carlo simulations using PRESTA EGS4 were employed to investigate the resulting dose enhancements from 4, 6, and 10 MV energies commonly used in the stereotactic radiosurgery system. Dose enhancements from 32% to 68% were observed close to the platinum implant for the above energies when using a 12.5 mm collimator. Comparatively higher dose enhancements were observed when using smaller collimators. It was found that this dose enhancement increased with beam energy but decreased as beam size increased.
机译:当高能和低能致stra光子从线性加速器辐射高原子序数的材料时,会发生三个光子相互作用过程,即光电效应,康普顿效应和成对产生。由于光电效应和成对产生,XKnife放射外科治疗计划系统无法预测放射外科手术中铂植入物(具有高原子序数)附近目标的剂量增加。在目前的工作中,采用了使用PRESTA EGS4的蒙特卡罗模拟来研究立体定向放射外科系统中常用的4、6和10 MV能量产生的剂量增加。使用12.5 mm准直器时,对于上述能量,在接近铂植入物的位置,观察到剂量增加了32%至68%。当使用较小的准直器时,观察到相对较高的剂量增加。发现该剂量增加随着束能量而增加,但是随着束大小的增加而减小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号