首页> 外文期刊>Medical Physics >Calibration of a scintillation dosemeter for beta rays using an extrapolation ionization chamber.
【24h】

Calibration of a scintillation dosemeter for beta rays using an extrapolation ionization chamber.

机译:使用外推电离室对闪烁剂量计进行β射线校准。

获取原文
获取原文并翻译 | 示例
       

摘要

A scintillation dosemeter is calibrated for 90Sr/90Y beta rays from an ophthalmic applicator, using an extrapolation ionization chamber as a reference instrument. The calibration factor for the scintillation dosemeter agrees with that given by the manufacturer of the dosemeter within ca. 2%. The estimated overall uncertainty of the present calibration is ca. 6% (2 sd). A calibrated beta-ray ophthalmic applicator can be used as a reference source for further calibrations performed in the laboratory or in the hospital.
机译:使用外推电离室作为参考仪器,对来自眼科敷料器的90Sr / 90Yβ射线校准闪烁剂量计。闪烁剂量计的校准因数与剂量计制造商在大约10分钟内给出的校准因数一致。 2%。当前校准的估计总体不确定性约为。 6%(2标清)。经校准的β射线眼科敷料器可用作在实验室或医院进行的进一步校准的参考源。