Improving the sensitivity of high-frequency subharmonic imaging with coded excitation: a feasibility study.

机译:利用编码激励提高高频次谐波成像的灵敏度:一项可行性研究。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Subharmonic intravascular ultrasound imaging (S-IVUS) could visualize the adventitial vasa vasorum, but the high pressure threshold required to incite subharmonic behavior in an ultrasound contrast agent will compromise sensitivity-a trait that has hampered the clinical use of S-IVUS. The purpose of this study was to assess the feasibility of using coded-chirp excitations to improve the sensitivity and axial resolution of S-IVUS.The subharmonic response of Targestar-p(TM), a commercial microbubble ultrasound contrast agent (UCA), to coded-chirp (5%-20% fractional bandwidth) pulses and narrowband sine-burst (4% fractional bandwidth) pulses was assessed, first using computer simulations and then experimentally. Rectangular windowed excitation pulses with pulse durations ranging from 0.25 to 3 μs were used in all studies. All experimental studies were performed with a pair of transducers (20 MHz/10 MHz), both with diameter of 6.35 mm and focal length of 50 mm. The size distribution of the UCA was measured with a Casy(TM) Cell counter.The simulation predicted a pressure threshold that was an order of magnitude higher than that determined experimentally. However, all other predictions were consistent with the experimental observations. It was predicted that: (1) exciting the agent with chirps would produce stronger subharmonic response relative to those produced by sine-bursts; (2) increasing the fractional bandwidth of coded-chirp excitation would increase the sensitivity of subharmonic imaging; and (3) coded-chirp would increase axial resolution. The experimental results revealed that subharmonic-to-fundamental ratios obtained with chirps were 5.7 dB higher than those produced with sine-bursts of similar duration. The axial resolution achieved with 20% fractional bandwidth chirps was approximately twice that achieved with 4% fractional bandwidth sine-bursts.The coded-chirp method is a suitable excitation strategy for subharmonic IVUS imaging. At the 20 MHz transmission frequency and 20% fractional bandwidth, coded-chirp excitation appears to represent the ideal tradeoff between subharmonic strength and axial resolution.
机译:次谐波血管内超声成像(S-IVUS)可以可视化外膜脉管血管,但是在超声造影剂中激发次谐波行为所需的高压阈值将损害敏感性,这是阻碍S-IVUS临床应用的一个特征。这项研究的目的是评估使用编码线性调频激发来提高S-IVUS的灵敏度和轴向分辨率的可行性。商业微泡超声造影剂(UCA)Targestar-p(TM)对首先使用计算机模拟,然后通过实验评估了编码线性调频脉冲(分数带宽的5%-20%)和窄带正弦脉冲(分数带宽的4%)脉冲。在所有研究中均使用了脉冲持续时间为0.25至3μs的矩形窗口激励脉冲。所有实验研究均使用一对直径为6.35 mm,焦距为50 mm的换能器(20 MHz / 10 MHz)进行。使用CasyTM Cell计数器测量UCA的尺寸分布。模拟预测的压力阈值比实验确定的压力阈值高一个数量级。但是,所有其他预测与实验观察结果一致。预计:(1)与线性正弦爆发相比,用线性调频脉冲激发该物质会产生更强的次谐波响应; (2)增加编码线性调频激励的分数带宽将提高亚谐波成像的灵敏度; (3)编码线性调频将提高轴向分辨率。实验结果表明,chi发出的次谐波与基本比率比持续时间类似的正弦爆发产生的谐波高5.7 dB。分数带宽20%线性chi的轴向分辨率大约是分数带宽正弦突发4%所达到的轴向分辨率的两倍。编码线性调频方法是适合于次谐波IVUS成像的激励策略。在20 MHz的传输频率和20%的分数带宽下,编码线性调频激励似乎代表了次谐波强度与轴向分辨率之间的理想折衷。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号