A method to estimate the effect of deformable image registration uncertainties on daily dose mapping

机译:一种估计可变形图像配准不确定度对日剂量映射的影响的方法

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To develop a statistical sampling procedure for spatially-correlated uncertainties in deformable image registration and then use it to demonstrate their effect on daily dose mapping. Methods: Sequential daily CT studies are acquired to map anatomical variations prior to fractionated external beam radiotherapy. The CTs are deformably registered to the planning CT to obtain displacement vector fields (DVFs). The DVFs are used to accumulate the dose delivered each day onto the planning CT. Each DVF has spatially-correlated uncertainties associated with it. Principal components analysis (PCA) is applied to measured DVF error maps to produce decorrelated principal component modes of the errors. The modes are sampled independently and reconstructed to produce synthetic registration error maps. The synthetic error maps are convolved with dose mapped via deformable registration to model the resulting uncertainty in the dose mapping. The results are compared to the dose mapping uncertainty that would result from uncorrelated DVF errors that vary randomly from voxel to voxel. Results: The error sampling method is shown to produce synthetic DVF error maps that are statistically indistinguishable from the observed error maps. Spatially-correlated DVF uncertainties modeled by our procedure produce patterns of dose mapping error that are different from that due to randomly distributed uncertainties. Conclusions: Deformable image registration uncertainties have complex spatial distributions. The authors have developed and tested a method to decorrelate the spatial uncertainties and make statistical samples of highly correlated error maps. The sample error maps can be used to investigate the effect of DVF uncertainties on daily dose mapping via deformable image registration. An initial demonstration of this methodology shows that dose mapping uncertainties can be sensitive to spatial patterns in the DVF uncertainties.
机译:目的:为可变形图像配准中的空间相关不确定性开发一种统计采样程序,然后用它来证明其对每日剂量映射的影响。方法:进行连续的每日CT研究,以在分割外束放射治疗之前绘制解剖学变化图。这些CT可变形地对准计划CT,以获得位移矢量场(DVF)。 DVF用于累积每天输送到计划CT上的剂量。每个DVF都有与其相关的空间相关不确定性。主成分分析(PCA)应用于测得的DVF误差图,以产生与误差相关的去相关主成分模式。分别对模式进行采样并进行重构,以生成综合的配准误差图。合成误差图通过可变形配准与剂量图卷积,以对剂量图中的不确定性建模。将结果与不相关的DVF误差(体素随体素随机变化)导致的剂量映射不确定性进行比较。结果:显示了误差采样方法可生成合成DVF误差图,与观察到的误差图在统计上无法区分。通过我们的程序建模的与空间相关的DVF不确定性产生的剂量映射误差模式与由于随机分布的不确定性而不同。结论:可变形的图像配准不确定性具有复杂的空间分布。作者已经开发并测试了一种将空间不确定性去相关并制作高度相关的误差图的统计样本的方法。样本误差图可用于通过可变形图像配准来调查DVF不确定性对每日剂量图的影响。该方法的初步证明表明剂量映射不确定性可能对DVF不确定性中的空间模式敏感。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号