Single x-ray absorptiometry method for the quantitative mammographic measure of fibroglandular tissue volume.

机译:单一X射线吸收法用于乳腺纤维体积定量X线摄影测量。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: This study describes the design and characteristics of a highly accurate, precise, and automated single-energy method to quantify percent fibroglandular tissue volume (%FGV) and fibroglandular tissue volume (FGV) using digital screening mammography. METHODS: The method uses a breast tissue-equivalent phantom in the unused portion of the mammogram as a reference to estimate breast composition. The phantom is used to calculate breast thickness and composition for each image regardless of x-ray technique or the presence of paddle tilt. The phantom adheres to the top of the mammographic compression paddle and stays in place for both craniocaudal and mediolateral oblique screening views. We describe the automated method to identify the phantom and paddle orientation with a three-dimensional reconstruction least-squares technique. A series of test phantoms, with a breast thickness range of 0.5-8 cm and a %FGV of 0%-100%, were made to test the accuracy and precision of the technique. RESULTS: Using test phantoms, the estimated repeatability standard deviation equaled 2%, with a +/-2% accuracy for the entire thickness and density ranges. Without correction, paddle tilt was found to create large errors in the measured density values of up to 7%/mm difference from actual breast thickness. This new density measurement is stable over time, with no significant drifts in calibration noted during a four-month period. Comparisons of %FGV to mammographic percent density and left to right breast %FGV were highly correlated (r=0.83 and 0.94, respectively). CONCLUSIONS: An automated method for quantifying fibroglandular tissue volume has been developed. It exhibited good accuracy and precision for a broad range of breast thicknesses, paddle tilt angles, and %FGV values. Clinical testing showed high correlation to mammographic density and between left and right breasts.
机译:目的:本研究描述了一种高精度,精确和自动化的单能量方法的设计和特征,该方法使用数字化乳腺X线摄影术来量化纤维腺组织体积百分比(%FGV)和纤维腺组织体积百分比(FGV)。方法:该方法使用乳房X线照片未使用部分中的乳房组织等效体模作为估计乳房组成的参考。该体模用于计算每个图像的乳房厚度和成分,而与X射线技术或桨叶倾斜的存在无关。体模附着在乳腺钼靶压迫桨的顶部,并保持在适当位置,以进行颅尾和中外侧斜向筛查。我们描述了一种自动方法,用于通过三维重建最小二乘技术识别体模和桨的方向。制作了一系列的测试体模,乳房厚度范围为0.5-8 cm,%FGV为0%-100%,以测试该技术的准确性和精确性。结果:使用测试体模,估计的重复性标准偏差等于2%,在整个厚度和密度范围内的精度为+/- 2%。未经校正,发现桨叶倾斜会在与实际乳房厚度相差高达7%/ mm的测量密度值中产生较大误差。这种新的密度测量随着时间的推移是稳定的,在四个月的时间内校准没有明显的漂移。 %FGV与乳腺摄影百分比密度和左右乳房%FGV的比较高度相关(分别为r = 0.83和0.94)。结论:已经开发了一种自动的方法来量化纤维腺组织的体积。对于宽范围的胸部厚度,桨叶倾斜角和%FGV值,它都具有良好的精度和精确度。临床测试显示与乳房X线照片密度以及左右乳房之间高度相关。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号