A respiratory-gated treatment system for proton therapy.

机译:用于质子治疗的呼吸门控治疗系统。

获取原文
获取原文并翻译 | 示例
           

摘要

Proton therapy offers the potential for excellent dose conformity and reduction in integral dose. The superior dose distribution is, however, much more sensitive to changes in radiological depths along the beam path than for photon fields. Respiratory motion can cause such changes for treatments sites like lung, liver, and mediastinum and thus affect the proton dose distribution significantly. We have implemented and commissioned a respiratory-gated system for range-modulated treatment fields. The gating system was designed to ensure that each gate always contains complete modulation cycles so that for any beam segment the delivered dose has the planned depth-dose distribution. Measurements have been made to estimate the time delays for the various components of the system. The total delay between the actual motion and the beam on/off control is in the range of 65-195 ms. Time-resolved dose measurements and film tests were also conducted to examine the overall gating effect.
机译:质子疗法具有出色的剂量适应性和减少整体剂量的潜力。但是,优越的剂量分布比沿着光子场对沿束路径的放射深度变化更敏感。呼吸运动会引起诸如肺,肝和纵隔等治疗部位的这种变化,从而显着影响质子的剂量分布。我们已经实施并调试了用于距离调节治疗领域的呼吸门控系统。门控系统的设计可确保每个门始终包含完整的调制周期,以使对于任何光束段,所输送的剂量均具有计划的深度剂量分布。已进行测量以估计系统各个组件的时间延迟。实际运动与光束开/关控制之间的总延迟在65-195 ms的范围内。还进行了时间分辨的剂量测量和薄膜测试,以检查总体门控效果。