A novel method of radiochromic film dosimetry using a color scanner.

机译:一种使用彩色扫描仪进行放射性变色胶片剂量测定的新方法。

获取原文
获取原文并翻译 | 示例
       

摘要

A procedure that allows the improved extraction of the dose information based on the multicolor scanning of the radiochromic film is presented. The basic principle is the determination of the dose values from each color channel of the digital film image in RGB format by applying a nonlinear calibration function. The best estimate of the dose is then a weighted mean of the dose values derived from each color channel. The weighting factors are determined in such a way that the noise in the two-dimensional dose profile is at the minimum. The calculation of the weighting factors is presented; they are chosen to be proportional to the signal-to-noise ratio, Si/v(i)2, in all three color channels, i=red, green, or blue. The data reduction can be fully computerized, including the "cleaning" of the digital image from dust and scratches. It is highly reproducible, which is important for quality assurance, and easy to use. Our novel evaluation procedure combines the good response in the low dose range of the red color with the extended dose range of the blue and green color channels (response up to 10,000 Gy), making use of one single, steady evaluation function. Therefore, a smooth evaluation is possible in a wide dose range. For the interpretation of measurements with the radiochromic films the spatial inhomogeneity of the film's response to ionizing radiation is very important. Investigations on both film types, HS and MD55-2, as well as on the new EBT film have been carried out.
机译:提出了一种允许基于放射变色膜的多色扫描改进剂量信息提取的程序。基本原理是通过应用非线性校准功能从RGB格式的数字胶片图像的每个颜色通道确定剂量值。然后,对剂量的最佳估计是从每个颜色通道得出的剂量值的加权平均值。以使得二维剂量分布中的噪声最小的方式来确定加权因子。给出了加权因子的计算;在所有三个颜色通道(i =红色,绿色或蓝色)中,将它们选择为与信噪比Si / v(i)2成比例。数据缩减可以完全计算机化,包括“清除”灰尘和划痕的数字图像。它具有很高的可复制性,对于保证质量非常重要,并且易于使用。我们新颖的评估程序结合了单一稳定的评估功能,将红色的低剂量范围内的良好响应与蓝色和绿色通道的扩展剂量范围(响应高达10,000 Gy)结合在一起。因此,可以在较宽的剂量范围内进行平稳评估。为了解释放射性变色膜的测量值,膜对电离辐射的响应的空间不均匀性非常重要。已经对HS和MD55-2两种胶片类型以及新型EBT胶片进行了研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号